Quantity One User Guide
8-4
8.2 Step I. Select Application
To set the appropriate filter and other parameters for the type of object you
are imaging, click on the Select button under Select Application. The available
applications and their associated settings are listed in a tree that expands
from left to right.
Fig. 8-3. The application tree in the Fluor-S acquisition window.
First select your general application, then select the particular stain or
medium you are using. If you select the chemiluminescent application under
“Blotting,” you also have the option of selecting High Resolution or High
Sensitivity (see below).
When you select an application, the software automatically sets the
appropriate standard filter (520LP, 530DF60, 610LP, clear, or none), light type
(UV, white, or none), and light source (trans, epi, or neither) in the Fluor-S for
that particular application.
For applications involving trans illumination, you must also specify a scan
dimension (see below).
Your selection will be displayed below the Select button. To exit the tree
without selecting, press the ESC key.