RAD Data comm FLM-1 Guide de l'utilisateur Page 108

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 401
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 107
Quantity One User Guide
6-10
The first time you use the GS-710, you must select some calibration settings to
ensure that your calibration is accurate.
To set the automatic calibration settings, click on the More Options button in
the GS-710 acquisition window. This will open the Densitometer Options
dialog box.
Fig. 6-9. Densitometer Options dialog box.
6.5.a Editing the Step Tablet
Before you scan for the first time, you must enter the step tablet values for
your transmissive calibration strip into the Step Tablet Values form.
Note: In reflective scanning, the scanner is calibrated against the reflective step
tablet. However, you do not need to enter the values for this step tablet;
you can use the default values in the Step Tablet Values form.
Attached to the outside of your GS-710, you will find a copy of the
manufacturer’s printout of the diffuse density values for each segment of
your transmissive step tablet, as well as a serial number that identifies the
Vue de la page 107
1 2 ... 103 104 105 106 107 108 109 110 111 112 113 ... 400 401

Commentaires sur ces manuels

Pas de commentaire